DermaClip Instructions for Use

The DermaClip device is a needle-free alternative to sutures, staples and glues for the closure of lacerations, wounds, and surgical incisions.

APPLYING DERMACLIP IS AS SIMPLE AS “PICK, PLACE, PRESS FIRMLY & PULL”

Click below for your preferred materials and method of training:

If looking for DermaQuik (OTC) instructions, please click here.

Full instructions are set out in the Instructions for Use (IFU) that are included with every purchase of DermaClip and can be downloaded here.

Contraindications:

  1. The device should not be used on patients with known allergies to skin adhesives.
  2. Do not apply the device under excessive tension or on wounds that cannot be easily approximated, as skin
    shearing, blistering, or device release may result.
  3. Do not use on infected wounds unless specifically determined appropriate by the attending physician.
  4. Do not use in areas on wounds where the device cannot be fully adhered to the skin.
  5. Do not use within the hairline unless the area has been depilated before applying.
  • Note:  The DERMACLIP Extreme is not Suitable for use on Fragile or Damaged Skin. For individuals with these types of skin issues, utilize the standard dermaclip products.

CLEAN

  • DermaClip needs to be applied to CLEAN, DRY skin
  • Use normal protocols for cleaning the wound area
  • Ensure area is dry before applying devices

PICK

  • Remove the devices from the sterile packaging
  • Select a DermaClip, pinch the pull tabs towards the center, then lift it off of the backing paper
  • Make efforts not to touch the skin / tacky side of the adhesive to ensure the strongest adherence once applied

PLACE, PRESS FIRMLY

  • Center the device across the wound
  • Place each pad as close to the wound edge as possible without crossing into the wound
  • FIRM PRESSURE IS REQUIRED TO ACTIVATE THE PRESSURE SENSITIVE ADHESIVE. THIS WILL ENSURE THE DEVICE ADHERES TO THE SKIN WITHOUT PRESSURE, THE DEVICE COULD FALL OFF PREMATURELY
  • Place multiple devices 1 to 4 mm apart until the wound is covered

PULL

  • Once the length of the wound is covered, pull the tabs on each DermaClip in opposing directions to lock
  • A “click” is heard and felt when locking shut
  • The best closures are achieved when the tabs are pulled “Down & Out”: down towards the skin and in an outward direction.

    PULLING “UP” ON THE TABS STARTS TO WORK AGAINST THE ADHESIVE’S BOND WITH THE SKIN.
  • Make sure hear the “CLICK” is heard to assure the device is locked shut.

IMPORTANT:  There is a first resistance as the device is initially pulled closed that then releases. THIS IS NOT THE LOCKING OF THE DEVICE!

It is very important to the proper functioning of the device that the click occurs after the two edges of the device have made contact and then forced each other, and the attached skin, into an upward position. Only then will the device be closed and locked shut.

CUT TABS

  • For convenience and comfort, cut the excess pull tab
  • Ensure the device is locked before cutting pull tab
  • Cover the wound area with a dressing
  • Ensure any adhesive or tape used to cover devices does not adhere to the DermaClip devices

To provide a visual aid to support the warning on locking, the image below shows the different positions as the device is closed

When locked properly, the angled faces of the DermaClip device have made contact and then forced each other into an upward, everted position.

DermaClip Video Instructions

DermaClip IFU Downloads

Full Instructions for Use (IFU)

For Part Numbers 61001 / 62001 / 63001

For Part Numbers 61002 / 62002 /63002

For Part Numbers 61003 / 62003 / 63003

For Part Numbers 61004 / 62004 / 63004

For Part Numbers 61005 / 62005 / 63005

IFU Supplements

For all DermaClip Products

For all DermaClip Products

For all DermaClip Products

Patient Care Instructions & Handouts

For all DermaClip Products, English & Spanish

For all DermaClip Products

In-Service Request

This page and all related content are only meant as a supplement to the full Instructions for Use (IFU).